​Te​ledyne LABS​ HLf -Series syringe pumps conform to safety standards for use in UL Class I, Division 2, Groups A B C & D, T4 environments. The HLf syringe pumps give you the same accurate, predictable flow and pressure control while meeting the safety requirement. The hazardous location rating is achieved through internal design modifications including the use of brushless DC motors and non-contact switches. This approach eliminates the need for purge boxes or other additional safety devices. Wetted materials are compatible with most aqueous and organic liquids, corrosive solutions, heated substances, liquefied gases, viscous fluids, or slurries and pastes.​​


Capacity
Flow Range*
(mL/min)
Flow
Acc​uracy**
Pressure Range
(psi, bar)
Standard Pressure Range
Standard Plumbing Ports
Dimensions Continuous Flow Range (mL/min)
Wetted Materials
1000x
1015 mL
0.001–408
0.5% of Setpoint
10–2,000
0.7–137.9
0.5% FS
1/4” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.01–265
N, H, PTFE
500x
507 mL
0.001–204
0.5% of Setpoint
10–5000
0.7–345
0.5% FS
1/8” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.001–132
N, H, PTFE
500xv
High Viscosity​​
507 mL
0.001–204
0.5% of Setpoint
10–5000
0.7–345
0.5% FS
3/8” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.001–132
N, H, PTFE
260x
266 mL
0.001–107
0.5% of Setpoint
10–9,500
0.7–655
0.5% FS
1/8” Valco 39.8x10.7x18.4 in
101x27x47 cm
0.001–70 N, H, PTFE, G, T
65x
68 mL
0.00001–25
0.3% of Setpoint
10–20,000
0.7–1,390
0.1% FS
1/8” Valco
39.8x10.7x18.4 in
101x27x47 cm
0.00001–16 N, H, PTFE
  ​Wetted Materials: N=Nitronic 50, H=Hastelloy C-276, PTFE= Polytetrafluoroethylene, G=Gold, T=Titanium​


Capacity
Flow Range*
(mL/min)
Flow Accuracy**
Pressure Range
(psi, bar)
Standard
Pressure Range
Standard
​Plumbing Ports
Dimensions ​Continuous Flow Range (mL/min)
1000HLf
1015 mL
0.001–408
0.5% of Setpoint
10–2,000
0.7–137.9
0.5% FS
1/4” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.01–265
500HLf
507 mL
0.001–204
0.5% of Setpoint
10–3750
0.7–258.6
0.5% FS
1/8” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.001–132
260HLf
266 mL
0.001–107
0.5% of Setpoint
10–7,500
0.7–517
0.5% FS
1/8” Valco 39.8x10.7x18.4 in
101x27x47 cm
0.001–70
100HLf
102 mL
0.01 μL–60
0.3% of Setpoint
10–10,000
0.7–689.5
0.5% FS
1/8” Valco
39.8x10.7x18.4 in
101x27x47 cm
0.00001–16
​​* Maximum and minimum flows are dependent on optimizing your pump system. Consult a Teledyne ISCO Product Specialist to determine the best method for your application. For additional information, please consult the factory. Teledyne ISCO is continuously improving its products and reserves the right to change specifications without notice. All brand or product names mentioned herein are trademarks or registered trademarks of their respective holders.
** Flow rate accuracy are based on select conditions of fluid type, pressure and leakage rate.

​Frequently Asked Questions​

W​​hat is a syringe pump and its use?

A syringe is a mechanical device that uses a plunger to extract a liquid from one location and move it to another, such as for the application of mass spectrometry. In the laboratory, syringe pumps are used to distribute precisely measured liquids into scientific instruments for experimentation and production.

What are the different types of syringe pum​​ps?

Pumps can have a single syringe or multiple syringes. Primary types of pumps include:

  1. Infuse-only 
  2. Infuse and withdraw 
  3. High-pressure 
  4. Continuous flow independent channels: a dual syringe pump

What are the common prob​lems with syringe pumps?

Because syringe pumps have moving parts, those parts can wear out over time. Many are computer driven, and user error or software problems can occur. 

What is the me​​​dical name for syringe pumps?

Infusion pump.