Dissolution
|
CD AutoPlus & Autofill
CD AutoPlus Features
Handles surfactants and buffered media
Sample timing options: At Point or Midpoint
Seamless integration with space-saving AutoFill™ fraction collector stacked on top
Convenient auto-start feature
Programmable end-of-test wash routine automatically prepares system for next test
Print to serial and network printers
21 CFR Part 11 security compliant with event logs, error logs, and full audit trail
USP / US FDA / ASTM / EP / JP / CE / CSA / RoHS compliant
AutoFill Features
Precision fraction collector fully compatible with all AutoPlus™ autosampler models
Wash port with end-of-test wash routine
Sample retrieval return to vessel accessory
CE / CSA / RoHS compliant
Choice of easy install collection racks with rack detection sensors
Compatible with 16 x 100 mm test tubes with evaporation caps, and 12 x 32 mm vials with pre-cut septa
AutoPlus
Weight (dry)
21.3 kg (47.0 lbs.)
Dimensions (H x W x D)
46.8 cm (18.4 in.) x 38.1 cm (15.0 in.) x 66.0 cm (26.0 in.)
Power
Input: 24 VDC, 3.75 A, 90 W
Only use external power supply
AutoFill
Weight (dry)
14.5 kg (32.0 lbs.)
Dimensions (H x W x D)
40.1 cm (15.8 in.) x 28.0 cm (11.0 in.) x 47.3 cm (18.6 in.)
Sampling
Minimum sample interval:
1 min (5 min is normal)
Maximum sample interval:
999 hours, 59 min (999:59)
Sample time points:
50 max
Maximum sample volume:
90% of vial capacity
Collection
13 x 100 rack: 18 x 8 test tubes, 144 total
16 x 100 rack: 18 x 8 test tubes, 144 total
12 x 32 rack: 21 x 8 vials, 168 total
UPLC rack (12 x 32): 3 trays 6 x 8 vials, 144 total
Septum: Pre-cut type only.
Sampler Volume Accuracy
0.5 to 5.0 mL: ± 0.05
5.1 to 10.0 mL: ± 0.10
10.1 to 15.0 mL: ± 0.15
Power
Input: 24 VDC, 3.75 A, 90 W
Only use external power supply
External Power Supply
Input: 100 - 240 V~,50 - 60 Hz, 1.5 A max
O
utput: 24 VDC, 3.75 A, 90 W
For more information, please contact your local Teledyne Hanson representative or email us at:
[email protected]