Automation
|
|
Sample Types
| Liquid samples, including drinking water and wastewater; Liquid samples containing up to 15 mm of sediment when measured from the bottom of an upright 40 mL vial
|
Sample Vials
| 84-positions for 40 mL VOA vials, single hole cap with PTFE-faced silicone septum, per USEPA specifications
|
Automation
| XYZ robotics with septum piercing needle
|
General Specifications
| |
Dimensions (H x W x D)
| 25.3” x 21.6”x 26.3” (64.2 cm x 54.8 cm x 66.8 cm)
|
Weight
| Unit weight: 70 lbs (31.75 kg)
|
Power Requirements
| 100-240VAC +/- 10%, 50/60 Hz, 2.5A, 300W
|
Environmental Specifications
| Operating Temperature: 10° to 30 °C; Storage Temperature: - 20° to 60 °C; Relative Humidity: 10% to 90% ; the
front cover and autosampler is corrosion resistant to waters with a pH range of 1 to 10
|
Gas Handling
| |
Sample Pathway
| 1/16” and 1/8” O.D. PEEK tubing; 1/8” PTFE Tubing
|
Gas Supply
| Ultra-high Purity (99.999%) pure Helium or Nitrogen; Incoming Pressure: 60-100 psi, (100 psi max)
|
Liquid Handling
| |
Sample Precision
| < 1% RSD (n=7 @ 5 mL delivery volume measured by weight)
|
Sample Pathway
| Glass, PEEK, EPDM and Ultem® for solenoid valve manifold. 1/16” (0.16 cm) OD PEEK tubing for liquid transfer
|
Water Supply
| Requires use of a pressurized blank water reservoir (included) |
Cleaning
| The entire liquid sample pathway can be rinsed using the high-temperature DI water rinse cleaning technique;
user defined rinse volume and number of rinses for the needle and glassware
|
Blanks
| Automatic blanks can be pulled from the blank water reservoir and spiked with standard/ surrogate allowing all
autosampler positions to be used for samples
|
Vial Cooling
| Cools sample tray to 4 °C or less as defined by most USEPA Methodologies (requires an external recirculating
cooling bath)
|
Standard Injection
| |
Standard Injection System
| Two standard injection systems utilizing 2-way dosing valves mounted on the valve manifold
|
Capacity
| Up to 20 μL in 1, 2, 5, 10 and 20 μL increments at 1 µL per 1 µL injection
|
Precision and Accuracy
| < 10% RSD measured by GC/MS for Fluorobenzene and Bromofluorobenzene, (n=7), based on 5 µL
injection volume
|
Standard Vessels
| Two 15 mL standard vessels, UV-protected for added standard stability; Standard vessels sealed
under pressure for standard concentration integrity
|
Heater
| |
Water Heater
| Variable Heat Control from 40 °C to 90 °C.
|
Service
|
|
Electronic Leak Check
| Ability to leak check the sample pathway of the system via an automated system leak check process.
|
Benchmark Test
| The system has a mode that will allow for full electromechanical testing including; valving, heater, liquid delivery
system, inputs and outputs
|
Diagnostics
| The system offers independent control of all valves and liquid handling mechanisms for troubleshooting
|