​Te​ledyne LABS​ HLf -Series syringe pumps conform to safety standards for use in UL Class I, Division 2, Groups A B C & D, T4 environments. The HLf syringe pumps give you the same accurate, predictable flow and pressure control while meeting the safety requirement. The hazardous location rating is achieved through internal design modifications including the use of brushless DC motors and non-contact switches. This approach eliminates the need for purge boxes or other additional safety devices. Wetted materials are compatible with most aqueous and organic liquids, corrosive solutions, heated substances, liquefied gases, viscous fluids, or slurries and pastes.​​


Capacity
Flow Range*
(mL/min)
Flow
Acc​uracy**
Pressure Range
(psi, bar)
Standard Pressure Range
Standard Plumbing Ports
Dimensions Continuous Flow Range (mL/min)
Wetted Materials
1000x
1015 mL
0.001–408
0.5% of Setpoint
10–2,000
0.7–137.9
0.5% FS
1/4” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.01–265
N, H, PTFE
500x
507 mL
0.001–204
0.5% of Setpoint
10–5000
0.7–345
0.5% FS
1/8” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.001–132
N, H, PTFE
500xv
High Viscosity​​
507 mL
0.001–204
0.5% of Setpoint
10–5000
0.7–345
0.5% FS
3/8” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.001–132
N, H, PTFE
260x
266 mL
0.001–107
0.5% of Setpoint
10–9,500
0.7–655
0.5% FS
1/8” Valco 39.8x10.7x18.4 in
101x27x47 cm
0.001–70 N, H, PTFE, G, T
65x
68 mL
0.00001–25
0.3% of Setpoint
10–20,000
0.7–1,390
0.1% FS
1/8” Valco
39.8x10.7x18.4 in
101x27x47 cm
0.00001–16 N, H, PTFE
  ​Wetted Materials: N=Nitronic 50, H=Hastelloy C-276, PTFE= Polytetrafluoroethylene, G=Gold, T=Titanium​


Capacity
Flow Range*
(mL/min)
Flow Accuracy**
Pressure Range
(psi, bar)
Standard
Pressure Range
Standard
​Plumbing Ports
Dimensions ​Continuous Flow Range (mL/min)
1000HLf
1015 mL
0.001–408
0.5% of Setpoint
10–2,000
0.7–137.9
0.5% FS
1/4” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.01–265
500HLf
507 mL
0.001–204
0.5% of Setpoint
10–3750
0.7–258.6
0.5% FS
1/8” NPT 40.3x10.7x18.4 in
102x27x47 cm
0.001–132
260HLf
266 mL
0.001–107
0.5% of Setpoint
10–7,500
0.7–517
0.5% FS
1/8” Valco 39.8x10.7x18.4 in
101x27x47 cm
0.001–70
100HLf
102 mL
0.01 μL–60
0.3% of Setpoint
10–10,000
0.7–689.5
0.5% FS
1/8” Valco
39.8x10.7x18.4 in
101x27x47 cm
0.00001–16
​​* Maximum and minimum flows are dependent on optimizing your pump system. Consult a Teledyne ISCO Product Specialist to determine the best method for your application. For additional information, please consult the factory. Teledyne ISCO is continuously improving its products and reserves the right to change specifications without notice. All brand or product names mentioned herein are trademarks or registered trademarks of their respective holders.
** Flow rate accuracy are based on select conditions of fluid type, pressure and leakage rate.

Syringe pumps are a cornerstone of modern laboratory automation, delivering precise fluid control across analytical chemistry, research, and industrial applications. These instruments automate the precise movement of liquids through controlled pressure and flow rates, eliminating manual handling errors, improving safety, and ensuring accurate, reproducible results.

Laboratory App​​lications and Uses

In analytical laboratories, syringe pumps serve as critical components in chromatography systems, sample preparation workflows, and reaction monitoring. Their precision enables exact reagent addition, controlled sample injection, and accurate dilution series - essential processes in method development and quality control testing.

For research applications, these instruments provide the exacting control needed for core analysis, polymer research, and microfluidic studies. The ability to maintain consistent flow rates makes them invaluable for applications like deep-well core analysis and geological simulations.

In industrial settings, syringe pumps automate critical processes in pharmaceutical manufacturing, chemical production, and materials testing. Their precise control capabilities ensure consistent product quality while reducing operator intervention and potential contamination risks.

Key Advantages of Modern Syringe Pump Technology

From basic fluid transfer to complex multi-step processes, modern syringe pumps provide three essential advantages that make them indispensable in laboratory settings: automation and control, reliable performance across applications, and enhanced laboratory efficiency.

  • Automation and Control: Teledyne LABS syringe pumps offer unprecedented control over fluid movement, with programmable flow rates ranging from sub-microliter to hundreds of milliliters per minute. Their versatile operating modes include constant flow, constant pressure, and external control capabilities, while digital interfaces enable seamless integration with existing laboratory systems. Advanced models feature automated refill capabilities, enabling continuous operation for extended processes.
  • Performance and Reliability: These systems deliver exceptional precision with flow accuracy up to 0.3% of setpoint for critical applications, maintaining stable performance across a wide pressure range. Their zero-pulse flow design ensures smooth delivery for sensitive analytical procedures, while material compatibility extends from aqueous solutions to highly viscous materials, providing versatility across diverse laboratory applications.
  • Laboratory Efficiency: By automating fluid handling processes, syringe pumps significantly reduce manual intervention requirements while minimizing valuable reagent waste through precise dispensing. Their capability for extended unattended operation allows laboratories to maximize throughput and productivity, while integrated safety features protect both personnel and sensitive laboratory processes, ensuring reliable operation even in demanding environments.

These instruments represent a crucial advancement in laboratory automation, supporting the increasing demands for precision, reproducibility, and efficiency in modern analytical workflows. Their implementation directly addresses key laboratory challenges: enhancing productivity, eliminating manual errors, and ensuring consistent results across applications.​seo

Frequently Asked Questions

What are the different types of syringe pum ps?

Pumps can have a single syringe or multiple syringes. Primary types of pumps include:

  1. Infuse-only 
  2. Infuse and withdraw 
  3. High-pressure 
  4. Continuous flow independent channels: a dual syringe pump

When should you use a syringe pump?

You should use a syringe pump when precise control over fluid delivery is necessary, such as administering medications at a specific rate in clinical settings. It is also indispensable during experimental procedures in research laboratories where the accurate dosing of reagents or solvents can impact the outcome of experiments. Moreover, when treatments necessitate continuous or programmed infusion of substances, syringe pumps are pivotal in achieving desired therapeutic outcomes with minimal risk.

How user-friendly is the pump interface?

Our versatile pumps offer unparalleled ease of operation through an intuitive screen interface. Users can quickly access all functions through the main menu, including flow rate adjustment, pressure monitoring, and programming options. The digital display provides real-time feedback and clear visibility of operating parameters.

Can these pumps handle complex chemical structures and sensitive materials?

Yes, our syringe pumps are designed to handle a wide range of materials, from simple aqueous solutions to complex chemical structures. The wetted materials are carefully selected to ensure compatibility with most laboratory chemicals, including corrosive solutions and organic solvents.​