The Fusion Total Organic Carbon (TOC) Analyzer and TekLink™ Software offer solutions for increased sample throughput while improving the reliability of the analysis data. A few of the features that provide increased ease of use are:

  • Unattended auto-calibration monitoring
  • Automatic dilution capabilities from stock standards or samples
  • Intellidilution for automatically diluting a sample back into calibration range.
  • Automated self-diagnostic tools including leak check
  • 21 CFR 11 functionality - Electronic Records and Electronic Signatures (ERES) including audit trails and user-account management
  • Exportable reports in user-definable formats including metadata and versioning
  • Turn-key method development and schedule builder​
  • Mass Flow Controller (MFC) for reduced gas consumption and enhanced flow control​​

Me​​thods

  • EPA 415.1- 415.3, 9060A
  • Standard Method 5310C, ASTM D4779 and D4839
  • prENV 13370
  • Cleani​ng Validation​ / USP TOC Method <643> / EP 2.2.44 / JP​

​Chemistry
​Photochemical Oxidation via UV-Persulfate
​Detector
​Nondispersive Infrared (NDIR) with Static Pressure Concentration (SPC) - Patented
​Analytical Modes
​TOC (NPOC), TC- IC, TC, IC
​Analytical
​Limit of Detection: 0.2 ppb Maximum Measurable Concentration: 4,000 ppm (sample volume and dilution dependent) Carryover: = 1.0% Cross Contamination Precision*: = 1.0% RSD,+/-2 ppb or +/- 0.02 µgC, typical of a mid-range standard (Whichever is greater over seven replicates). * Analytical performance affected by laboratory water, reagent and gas purity, as well as sample container cleanliness, sample matrix, gas regulator cleanliness and precision, and operator skill.
​Complete Process and Analysis Time
​4-8 minutes typical for TOC analysis; Typically 12-22 minutes for Triplicate TOC Analysis
​Controller
​PC, Interface through Windows® 7 Professional or greater
​21 CFR Part 11 Software Control
​TOC TeklinkTM Software is a 21 CFR Part 11 tool for your laboratory compliance
​Data Handling
​​• Pre-defined Industry Standard Methods and Customized User Defined Methods
• Priority samples via schedule interrupt
• Real-time and Historical graphical display of NDIR detector data
• Reports exportable XML, CSV and HTML format
• Recalculation of data, outlier deletions, and precision performance criteria controls
• Ability to view historical results from multiple schedules on one graphical display
​Calibration​​
​• Auto-Calibration from Single Stock Standards or User Calibration Standards
• Multi-point (Linear or Quadratic) and auto-blanking
• Ability to use one calibration curve and blank for entire instruments’ analytical range
• Auto-Check Standards from Single Stock Standards or User Calibration Standards
  - Pass / Fail Criteria
  - Decision Control upon Failure (Halt, Re-Calibrate, or Continue)
​Other Features
​• Auto-System Suitability with Performance Measurements
• Auto-dilution of samples/standards
• Validation Support Package Available
• Pre-programmed point and click method setup
• Programmable flow rate and pressure control and monitoring
​• Auto-Leak Check
• Automatic shutdown/standby
• Self-cleaning sample handling process that cleans reactor chambers on every repetition
​• Intellidilution
​Official Methods
​EPA 415.1- 415.3, 9060A, Standard Method 5310C, ASTM D4779 and D4839, and prENV 13370, Cleaning Validation / USP TOC Method <643> / EP 2.2.44 / JP
​Dimensions
​18 inches (45.7 cm) W x 24.5 inches (62.2 cm) D x 32 inches (81.3 cm) H
​Carrier Gas Supply
​99.99% pure nitrogen cylinder or 99.9% pure nitrogen generator
​Inlet Carrier Gas Pressure
​65 to 100 psi​

Windows® is a registered trademark of Microsoft, TekLink™ is a registered trademark of Teledyne Tekmar Company. 
Covered by one or more of the following patents: 7,651,866; 8,128,874.
1UV Lamp contains Mercury, Do Not Put In Trash. Recycle or Dispose as Hazardous Waste.​