The triple mode, no enrichment, single or double gold amalgamation QuickTrace M-8000 CVAF is the system of choice for labs wanting to meet the most stringent reporting limits for academic research or enhanced waste discharge limits demanded by the “Clean Water Act”.
- Non-foaming Gas Liquid Separator (GLS) with overflow prevention system
- Automatic end of run and inactivity standby routines
- ≤ 0.05 ng/L instrument detection limits
- Advanced contamination control, over range and smart rinse features
- Intuitive gas controls eliminating air infusion into the system during sample probe movements during mode 2 and 3 analysis
- EPA 1631 and EPA 245.7 compliant (gold amalgamation modes)
- “Smart Rack” technology
- 12-roller 4-channel peristaltic pump
- Sample volume 0.5 mL to > 50 mL
- Unlimited QC sample positions
Performance
The QuickTrace M-8000 CVAF analyzer is designed for exceptional short and long-term stability. Using 25 mL or less of sample, an Instrument Detection Limit (IDL) less than 0.02 ng/L is typical using gold trap modes. When consuming less than 10 mL of sample, the M-8000 consistently achieves an IDL less than 0.1 ng/L in non-gold trap mode.
- Ultra-trace Instrument Detection Limit (IDL) (≤0.05 ng/L)
- Linearity greater than four orders of magnitude
- Dynamic range <0.05 ng/L to >400 µg/L
- In Mode 1, by response comparison to a 100 ng/L standard, the system exhibits <0.01% memory effect of a 1 mg/L standard, immediately following the ingestion of a 1 part per million (mg/L) sample
- In Mode 1, without employing the system's smart rinse technology, the analyzer is ready to accurately measure a sample within four minutes following the ingestion of a 1 ppm sample
- In Mode 2 and 3 - short-term precision (%RSD @ 95% Confidence) = 2.5% @ 5 ng/L, n=5
QuickTrace® Software
Teledyne LABS has designed the multi-tasking QuickTrace® software package to be easy to learn, yet provide valuable features and flexibility while continuously adding customer requested features, truly making the software designed by the customer for the customer.
Learn More >
Sample Automation
Customers have the freedom to choose an autosampler for their mercury analyzer that will meet their laboratories sample load requirements. The ASX-280 and ASX-560 autosamplers are available for any QuickTrace™ Mercury Analyzers. Manual sampling is also available with the option to add autosamplers as sample demand increases. See Specifications for details.
Methods
- EPA Method 245.7
- EPA Method 1631
- SW 846-7474
- EU Methods: Din ISO 16772
- EN-12338
- ISO 17852
*For a full list of applicable methods contact your sales specialist.