Simple Operation​​

Operational panel p​rovides easy-to-use controls. One simple button initiates complete media measurement/deaeration/dispense cycle. Fills dissolution vessels in less than 5 minutes.​

finger pushing button on control panel​​​​​​​

Volume Adjustment

Volume change adjustments provide quick and easy volume settings from 250 mL to 1,000 mL in 50 mL increments. Fill 6 or 7 vessels in Hanson, Agilent, VanKel, Distek, Erweka, or other dissolution testers.​

graduated volume measurement device 

Media-Mate Plus™ Features

  • Preheats, filters, deaerates, and volumetrically dispenses media in less than 5 minutes
  • Delivers media to 6 or 7 vessels simultaneously
  • Easy adjustments for precise volumetric dispensing with ±1% accuracy
  • Mobile, ergonomic cart for easy use around the lab
  • Convenient retractable power cord
  • Measuring vessels made with corrosion-resistant parts
  • Compatible with nearly all dissolution testers
  • Interchangeable pre-heat tanks (10-40 L) can be conveniently dedicated to different media and buffers
  • USP <711> and <1092> compliant
  • Full compliance to USP Toolkit (dissolved oxygen below 6 ppm)
  • The most important tool for ensuring successful PVT tests​
  • Interchangeable pre-heat tanks (10-40 L) can be conveniently dedicated to different media and buffers. Order an extra tank, heater, and temperature control for this purpose.
Aut​oPlus

​Weight
​Without media: 67 kg (148 lbs.)
With 40 liters of media: 107 kg (236 lbs.) 
Dimensions (H x W x D)
114.0 cm (44.8 in.) x 53.4 cm (21.0 in.) x 80.9 cm (31.9 in.)​
Electrical
Voltage: 115 VAC or 230 VAC ± 10%
Frequency: 50/60 Hz
Current: 13 Amp (115 VAC) or 8 Amp (230 VAC)
Phase: single
Environment
​Maximum ambient temperature: 5 °C below temperature set point
Maximum humidity: < 85 %
​Performance​​
​ ​​
​​​​Cycle time​
​less than 5 minutes​
Pre-heat temperature range
30 °C to 45 °C ​
Pre-heat temperature control accuracy
​± 0.5 °C​ ​
Pre-heat temperature readout accuracy
​± 0.2 °C
​Pre-heat time (21 to 37 deg C)
​60 minutes
​Delivery volume range
​250 mL to 1000 mL 
​Delivery volume accuracy
​± 1.0 % 
Deaeration
​95% saturation maximum 
Filtration
​5 micron nominal 
​Tank capacities​
​Media tank volume (usable)
​40 liters (10.5 gal) 
​Media tank volume (unusable)
 2 liters (0.53 gal)
Media volume to top of filter
​8 liters (2.1 gal)
Low water level sensor volume
​~18 liters (4.8 gal) ​
Wetted Materials​
Tank
​High density polyethylene 
Heater
​Teflon coated 
Filter
​Polyethylene
Vessel
​SAN (styrene/acrylonitrile) 
Fittings and fill hose​
PVC
Tubing 
Stainless steel and Tygon 


Media-Mate Plus​​

Media-Mate Plus, 10-40 L, 115 V 
25-710-101
Media-Mate Plus, 10-40 L, 230 V 
​25-710-102
Q-Pak™ Self-Validation Guidelines for Media-Mate Plus ​
25-710-803
​Accessories​​
​ ​​
​Tank, Heater, and Temperature Control​, 115 V
​25-710-115
Tank, Heater, and Temperature Control, 230 V
25-710-116
​Temperature Control only, 115 V
25-710-125
​Temperature Control only, 230 V
25-710-126
​Heater Assembly only, 115 V
25-710-130
​​​Heater Assembly only, 230 V
25-710-131
​Single Vessel Filler
25-710-415
​Media Filler Hose Extension w/ Filter
11-000-004
​Filter Cartridge for Tank, Polypropylene, 5 micron, (5/pk)
25-710-723
​Low-Volume Tank Assembly
25-710-442
​Low-Volume Tank Upgrade Kit
25-710-731
​Fill Tube Upgrade Kit
25-710-733
Dispense Bar Assembly for Sotax AT7
25-710-113
​Calibration Gauge Set (required for fine volume adjustment)
91-902-045
​Calibration Kit (includes 8 Plastic Vessels (1 L), 8-Position Vessel Carrier, and Calibration Gauge Set)
25-710-705
​Base Plate Replacement Kit
​25-710-726

For more information, please contact your local Teledyne Hanson representative or email us at: