Automation
|
|
Sample Capacity:
|
84-positions for 40 mL VOA vials.
|
Vial Size:
|
Nominal 40 mL capacity, single hole cap with PTFE-faced silicone septum, per EPA specifications; 3 3/4“ high without cap and septum; 1 1/16” OD; 24 mm ID cap for water sampling.
|
Sample Handling
|
|
Liquid Handling:
|
Sample syringe (25 mL) dispenses variable volumes of water from 1 - 25 mL in 0.1 mL increments. 1/16” OD PEEK™ tubing for liquid transfer
|
Sample Gas Pathway:
|
Glass, PEEK™, SilcoNert® 2000 and PTFE for sample handling.
|
Cleaning:
|
The entire liquid sample pathway can be rinsed using a combination of the methanol rinse and the high-temperature DI water rinse cleaning techniques. User defined rinse volume and number of rinses for the needle and glassware.
|
Gas Handling
|
|
Electronic MFC:
|
System is capable of controlling flow rates between 5-500 mL/min variable between each mode of operation.
|
Pressure Monitoring:
|
Ability to record purge and bake pressure for each sample.
|
Gas Supply:
|
Ultra-high purity (99.999%) Helium or Nitrogen; Incoming Pressure: 65 - 100 psig, (100 psig max) |
Standard Injection
|
|
Standard Injection:
|
Three standard injection systems utilizing 2-way dosing valves mounted on an internal valve manifold
|
Capacity:
|
1, 2, 5, 10, and 20 µL increments.
|
Consumption:
|
1 µL per 1 µL injection
|
Standard Vessels:
|
Three 15 mL standard vessels, UV-protected and sealed under pressure for standard concentration integrity
|
Liquid Samples
|
includes drinking water and wastewater; Liquid samples containing up to 15 mm of sediment when measured from the bottom of an upright 40 mL vial
|
Sample Glassware:
|
The system is capable of operation with 5-25 mL frit or fritless sparge vessels. Ships standard with 5 mL fritted glassware.
|
Sample Dilutions:
|
Programmable automatic aqueous sample dilutions of 1:100, 1:50, 1:25, 1:10, 1:5, 1:2
|
Blanks:
|
Automatic blanks can be pulled from the water reservoir and spiked with standard allowing all autosampler positions to be used for samples.
|
Cycle Time:
|
Total Purge & Trap cycle time of 18 minutes, or less depending on the method.
|
Low-Level Solid Samples
|
includes all types of natural soils and sediments. Sampled: Direct purge in the vial per USEPA 5035 low-level soil methodology
|
Sample Needle:
|
A 3-stage needle allows for water and standards to be directly added to the vial where the solid sample will be purged.
|
Vial Heater:
|
Variable heat control from ambient to 90 °C.
|
Mixing:
|
The solid sample can be mixed via a stir bar using three variable speeds.
|
High-Level Solid Samples
|
includes all types of soils and sediments. Sampled: Automated Methanol Extraction and subsequent dilution per USEPA 5035 high-level soil methodology
|
Extraction:
|
System is capable of fully automating the methanol extraction of high-concentration soil samples.
|
Matrix Spike:
|
The system is configured to allow a surrogate spike to be added directly to the solid sample when the methanol is added for the extraction.
|
Extraction Dilutions:
|
Programmable automatic dilutions of methanolic extract of 1:100 or 1:50 for 5 mL sample volumes
|
System Control
|
|
Instrument Control:
|
Atomx XYZ TekLink™ software in a Windows® 7 or greater environment. Via USB
|
Method Scheduling
|
All method types can be run from any position in the sample sequence. Up to three standards can be added to any user-specified position. Multiple runs can be made from the same vial (not recommended).
|
System History:
|
The system records a complete history of all sample, schedule and method information.
|
Service
|
|
Electronic Leak Check:
|
Ability to leak check the entire sample pathway of the system via the automated System Leak Check. Built-in diagnostics that, once a leak has been identified, the system will check independent sub-systems for leaks.
|
Benchmark Test:
|
The system has a mode that will allow for full electromechanical testing including; valving, heaters, vial handling systems, liquid delivery system, inputs and outputs.
|
Diagnostics:
|
The system offers independent control of all valves, vial handling mechanisms and syringe drive for troubleshooting.
|
General Specifications
|
|
Dimensions & Weight:
|
Dimensions: 70.1 cm (27.6” W) X 58.42 cm (23” D) X 49.5 cm (19.5” H)
Unit weight: 43.1 kg (95 lbs
|
Power Requirements:
|
100-120VAC (±10%), 50/60 Hz, 10.0A, 1150W; 220-240VAC (±10%), 50/60 Hz, 5.0A, 1150W
|
Environmental Specs:
|
Operating Temperature: 10°-30 °C; Storage Temperature: -20°-60 °C; Relative Humidity: 10%-90%
|
Corrosion Resistance:
|
The front cover and sample tray are corrosion resistant to waters with a pH range of 1-10.
|