Automation
|
|
Sample Types
|
Liquid samples, including drinking water and wastewater; Liquid
samples containing up to 15 mm of sediment when measured from the bottom of an
upright 40 mL vial
|
Sample Vials
|
84-positions for 40 mL VOA vials, single hole cap with PTFE-faced
silicone septum, per USEPA specifications
|
Automation
|
XYZ robotics with septum piercing needle
|
General Specifications
|
|
Dimensions (H x W x D)
|
25.3 x 21.6 x 26.3 (64.2 cm x 54.8 cm x 66.8 cm)
|
Weight
|
Unit weight: 70 lbs (31.75 kg)
|
Power Requirements
|
100-240VAC +/- 10%, 50/60 Hz, 2.5A, 300W
|
Environmental Specifications
|
Operating Temperature: 10 to 30 C; Storage Temperature: - 20
to 60 C; Relative Humidity: 10% to 90% ; the front cover and autosampler is
corrosion resistant to waters with a pH range of 1 to 10
|
Gas Handling
|
|
Sample Pathway
|
1/16 and 1/8 O.D. PEEK tubing; 1/8 PTFE Tubing
|
Gas Supply
|
Ultra-high Purity (99.999%) pure Helium or Nitrogen; Incoming
Pressure: 60-100 psi, (100 psi max)
|
Liquid Handling
|
|
Sample Precision
|
< 1% RSD (n=7 @ 5 mL delivery volume measured by weight)
|
Sample Pathway
|
Glass, PEEK, EPDM and Ultem for solenoid valve manifold. 1/16
(0.16 cm) OD PEEK tubing for liquid transfer
|
Water Supply
|
Requires
use of a pressurized blank water reservoir (included)
|
Cleaning
|
The entire liquid sample pathway can be rinsed using the
high-temperature DI water rinse cleaning technique; user defined rinse volume
and number of rinses for the needle and glassware
|
Blanks
|
Automatic blanks can be pulled from the blank water reservoir and
spiked with standard/ surrogate allowing all autosampler positions to be used
for samples
|
Vial Cooling
|
Cools sample tray to 4 C or less as defined by most USEPA
Methodologies (requires an external recirculating cooling bath)
|
Standard Injection
|
|
Standard Injection System
|
Two standard injection systems utilizing 2-way dosing valves
mounted on the valve manifold
|
Capacity
|
Up to 20 μL in 1, 2, 5, 10 and 20 μL increments at 1 L
per 1 L injection
|
Precision and Accuracy
|
< 10% RSD measured by GC/MS for Fluorobenzene and
Bromofluorobenzene, (n=7), based on 5 L injection volume
|
Standard Vessels
|
Two 15 mL standard vessels, UV-protected for added standard
stability; Standard vessels sealed under pressure for standard concentration
integrity
|
Heater
|
|
Water Heater
|
Variable Heat Control from 40 C to 90 C.
|
Service
|
|
Electronic Leak Check
|
Ability to leak check the sample pathway of the system via an
automated system leak check process.
|
Benchmark Test
|
The system has a mode that will allow for full electromechanical
testing including; valving, heater, liquid delivery system, inputs and
outputs
|
Diagnostics
|
The system offers independent control of all valves and liquid
handling mechanisms for troubleshooting
|