Get the Peaks You Want

Collect only the peaks you want by programming desired peak detector optio​ns.​​

Level Sens​​ing​​​
slope with horizontal line

Collect peaks above detector threshold, enabling “heartcut” collection of concentrated part of the peak.​

Slope​
slope with vertical bounds

Collect peaks based on inflection points. Separate close peaks by valley triggers.

Time​ Window
clock with time wedge

Collect only during a specific interval of time.

Del​ay Compensation
hour glass

Cut peaks at the diverter valve regardless of system tubing lengths.​​​​​​​

​​​
Foxy​​ R1
​Inlet Tubing and Flow Ra​tes ​1/16" OD (0.031" ID), up to 25mL/min
(25 to 125 mL/min with high flow option)
Maximum Tube Height
​160 mm​
Racks
​1, no detection
​Maximum Tube Qty
​144 12/13 mm tubes
Construction Materials:
Racks:
Funnel Racks:
Diverter Valve:
Construction:
Drip Tray:
Drop Former:​

Stainless steel and polypropylene
ECTFE (Halar) or ETFE (Tefzel) funnels
PEEK and FFKM (fluoroelastomer)
Cast Aluminum
Polypropylene
PEEK
Solvent Compatibility
​Aqueous and Organic
Test Tube Time Change
​0.10 – 0.58 seconds (dependent on collection vessel tube advance speed)
Memory
​Eight methods​
Collection Modes
Time:
Drop:
Volume:

00:00:01– 99:59:59, 1 second intervals
1– 999 drops per fraction
Up to 999 units
​Peak Detection
​Slope, Threshold, Three time windows, Combination of any of the above collection modes
​​System Control
​Front panel touch screen, Ethernet or RS-232
Inputs
​Pump counter, external peak, inject, tube advance: +3.3 to 5 VDC or contact closure, active low, 30 ms minimum, 5 VDC max. Analog peak voltage: 0 to 1.0 V nominal, –1.2 to + 1.2 V usuable, –5 to +5 V max.
Outputs
​Pump pause/stop and event (tube change), 300 ms: open collector, active low, 15 VDC/100 mA maximum.
Automatic Shut-off
​Rack full, last tube, end of time window or external signal 
​Security Valve (2-way)
​n/a
Diverter Valve (3-way)
​Included
​Valve Pressure Rating
​0– 28 ps
​Dimensions (W x H x D)
​12.25 x 14.0 x 13.0 in
31.1 x 35.5 x 33.0 cm
Weight
​7.1 kg (15.6 lbs) 
Power Requirements
​100 – 240 VAC, 50–60 Hz, 1 ampere max.
Includes one IEC power cord supplied for regional AC power (100, 117, 234 VAC)
Certification
​CE, CSA